A mammogram is a special, low-dose X-ray technique used to take a picture of the breast, detecting and diagnosing any abnormal lumps or masses in breast tissue.


Features

  • 1. Unique adopt all-solid-state high frequency high voltage generator. This technology has got the PATENT IN THE USA.
  • 2. The safest mammography at high voltage. There is a built-in X-ray ignition coil in host machine, high-voltage power lines less then 25cm.
  • 3. Electric Isocentric rotating C-arm with a unique automatic back to center function.
  • 4. Optional the third generation imported moving grid.
  • 5. Optional auto/semi-auto/manual, three kind exposure modes.
  • 6. Optional image output device: digital film printer.
  • 7. A total of 3 pieces of large size full color LCD screen display, operation table 8 inch LCD screen is a touch key.
  • 8. Comfortable Compression
    When some degree of pressure is required for radiography, it allows you to presser the appropriate pressure(up to a maximum of 20kg)and is equipped with MICOM Control’s Soft-touch system which is designed to minimize the discomfort of the examine with in the pressure range.
    Tissue Compression: Manual and Motorized (Max 20kg)
    Compression Force and Thickness Data Display
    Micro Control’s Compression
    Automatic Release
  • 9. Intelligent Automatic Exposure Control(AEC)
    With the Automatic Exposure Control system, it is possible to produce images with reliable intensity suitable for and film, screen, or method of radiography.
    Furthermore, it greatly enhances the convenience of radiography by embedding the Full-AEC function which is capable of utilizing the Auto kV
    Type: Solid-State Detector
    Microprocessor Control
    AEC Mode: Full AEC(Auto kV)
    Semi AEC (kV Select)
    Manual (kV,mAs Select)
    Density Adjustment: 16 density steps
    10. X-ray Tube
    Model: China Hangzhou LR01
    Focal Spot Size: Dual Focus 0.2 / 0.4mm
    Target Material: Molybdenum (Mo)
    Port Material: Beryllium (Be)
    High-speed anode drive: 2800 /1000rpm
    Target angle:12°/12°
    Anode Heat Storage: 100KJ(150KHU)
    Anode Cooling: Air cooling
    Filtration: Mo(0.03mm)Al(0.5mm)

Specification

Item Parameter Remark
X-ray Generator Generator Type: High Frequency Inverter 80kHz Self-developed and world advanced  all-solid-state high frequency high voltage x-ray generator
Radiographic Ratings: Large Focal Point 20-35kV/10-510mAs
Small Focal Point 20-35kV/10-100mAs
Power Rating: 6kVA
X-ray Tube Focal Spot Size: Dual Focus 0.1 / 0.3mm Model: China Tube
Target Material: Molybdenum (Mo)
Port Material: Beryllium (Be)
High-speed anode drive: 2800 /10000rpm
Target angle:10°/16°
Anode Heat Storage: 300kHU
Anode Cooling: Air cooling
Filtration: Mo(0.03mm)Al(0.5mm)
Radiographic Stand C-ARM Vertical Movement: 590mm
Center of electric rotating C-arm, automatic return function by one key
Rotations Degree: +90°~-90°
Automatically released after the exposure pressure settings display. Compression flexible stepless speed.
Max. pressure: 200N
Max. travel: 150mm
SID: 650mm
Cassette Image Receptor Bucky Device: 18×24cm Bucky drive mechanism 24x30cm for optional
Grid Ratio: 5:1, 30 Line/cm
Others Line Voltage 110V for optional
220V ac±10%  25A, Single phase