CPAP ventilation modes keeps children spontaneous breath and reduces children breath work.


Product Technical Data

1)Parameter adjustment
CPAP adjusting range 0 cmH2O ~ 10 cmH2O
Continuous flow adjusting range 1.5 L/min ~ 15 L/min
Oxygen Concentration 21%-100%
2)Parameter monitoring range
CPAP/airway pressure
Monitoring range 0 cmH2O ~ 60 cmH2O
Accuracy ±(2 % full scale + 4 % actual value)
Continuous flow
Monitoring range 1.5 L/min ~ 15 L/min
Accuracy  ±4 %
3) Gas supply
Gas source  O2(For medical use), Air(For medical use)
Rated working pressure  350 kPa
Output pressure of regulator  350 kPa±50 kPa (300 kPa~400 kPa)

Product Working environment

Ambient temperature 5 ℃~ 40 ℃
Relative humidity  ≤80 %
Atmospheric pressure 960 hPa~1040 hPa
Storage and transportation environment
Ambient temperature -10 ℃~+40 ℃
Relative humidity ≤90 %
Atmospheric pressure 860 hPa~1060 hPa