Air temp. and skin temp. servo-controlled by computer.


Product Technical data

  • power supply: AC 110/220V; 60/50Hz
  • power input: 260W
  • operating environment:
    Ambient temperature: 10℃~30℃
    Relative humidity: 30%~75%
    Atmospheric pressure: 700hPa~1060hPa
    Air velocity of flow: ≤1m/s
  • Air temperature control range: 25℃~38℃
  • Skin temperature control range: 32℃~37.5℃
  • Air (or skin) temperature sensor accuracy: ±0.3℃
  • Temperature fluctuation: ±1℃
  • Uniformity of mattress temperature: ≤1.5℃
  • Air temperature deviation alarm: ±3.0℃
  • Skin temperature deviation alarm: ±1.0℃
  • Warm-up time: ≤40 min
  • Internal noise level: ≤60dB(A)
  • Main battary: DC 12V 40AH, Li-ion battery
  • Auxiliary battery: DC 8.4V 0.17AH, Li-ion battery
  • Air filter: 0.5μm

Transport and storage

  • ◎ Environment temperature: -40~+55℃
  • ◎ Environment humidity: ≤95%RH
  • ◎ Atmospheric pressure: 500~1060 hPa

Standard configration and optional parts

  • Main body (including the transparent double hood, control system, infant bed, mattress, skin temperature sensor, two oxygen cylinder, IV pole and lighting lamp)
  • Optional parts: Trolly (two specifications for option)

Product Package Information

  • Size (hood and controller): 1090×625×845 (mm), N.W.:60.5Kg; G.W: 110Kg
  • Trolley: 1405×650×940(mm) or 2080×640×470 (mm), N.W.:27Kg; G.W: 57Kg or N.W.:35Kg; G.W: 65Kg