CPAP ventilation modes keeps children spontaneous breath and reduces children breath work.
Product Technical Data
| 1)Parameter adjustment | |
| CPAP adjusting range | 0 cmH2O ~ 10 cmH2O |
| Continuous flow adjusting range | 1.5 L/min ~ 15 L/min |
| Oxygen Concentration | 21%-100% |
| 2)Parameter monitoring range | |
| CPAP/airway pressure | |
| Monitoring range | 0 cmH2O ~ 60 cmH2O |
| Accuracy | ±(2 % full scale + 4 % actual value) |
| Continuous flow | |
| Monitoring range | 1.5 L/min ~ 15 L/min |
| Accuracy | ±4 % |
| 3) Gas supply | |
| Gas source | O2(For medical use), Air(For medical use) |
| Rated working pressure | 350 kPa |
| Output pressure of regulator | 350 kPa±50 kPa (300 kPa~400 kPa) |
Product Working environment
| Ambient temperature | 5 ℃~ 40 ℃ |
| Relative humidity | ≤80 % |
| Atmospheric pressure | 960 hPa~1040 hPa |
| Storage and transportation environment | |
| Ambient temperature | -10 ℃~+40 ℃ |
| Relative humidity | ≤90 % |
| Atmospheric pressure | 860 hPa~1060 hPa |