Bipolar Cut Bipolar Coag fully meet different surgery requirement.


Application

Spinal Surgery, Obstetrics and Gynecology, ORL/ ENT surgery, Orthopedics & Plastic Surgery, Oral Surgery.

Technical Specification

Working  Mode Max Power Rated  Load
Monopolar Cut Pure cut 60W 500Ω
Blend 1 50W 500Ω
Monopolar Coag Standard Coag 50W 500Ω
Bipolar  Cut Bipolar Cut 40W 100Ω
Bipolar  Coag Bipolar Coag 40W 100Ω
Electrical requirement
Monopolar Measuring Frequency  4.0MHZ ±2% Bipolar Measuring Frequency 1.84MHZ ±2%
Max Input Currency 1400VA
  Environmental Conditions for Use
Condition Range
Temperature 50°C to 40°C
Relative humidity ≤ 95% non-condensing
Pressure 700 hPa to 1060 hPa
Environmental Conditions for Transport and Storage
Condition Range
Temperature -40°C to 55°C
Relative  humidity ≤ 93% non-condensing
Pressure 500 hPa to 1060 hPa
Dimension (L×W×H) 350mm×280 mm×138mm
Net Weight 8kg

Standard Accessories

Code Accessories Name Quantity
1 RF Electrosurgical Unit 1
2 Reusable ESU Pencil 1
3 Disposable Electrode 9
4 Reusable patient plate with cable 1
5 Resuable Bipolar forceps 1
6 Reusable Bipolar cable 1
7 Double  Footswitch 1
8 Power cable 1
9 English Service Manual 1