Portable and delicate design, easy to operate.


Product Technical Specification

Lead Standard 12 leads Time constant ≥3.2s
Lead acquisition 12bit/1000Hz (synchronously 12 leads) Frequency response 0.05~160Hz(-3db)
Work mode Man / Auto/Analysis/Storage Noise level <15μVp-p
Filter AC Filter : 50Hz / 60Hz

EMG Filter : 25Hz / 45Hz

Anti-Drift Filter : 0.15Hz (Adaptive)

Inter-channel Interference ≤0.5mm
CMMR ≥120dB (with AC FILTER) Sensitivity 5,10,20, 40, Auto  (mm/mV)
Input circuit Floating; Protection circuit against

Defibrillator effect

Recording mode 3ch+++, 6ch, 6ch+, 12ch

Default Man. Mode is 3ch+++ format

Default Auto. Mode is 12 format

Input Impedance ≥50MΩ Paper speed 6.25、12.5、25、50mm/s(±3%)
Input circuit current ≤0.05μA Paper 215mm *20m or 210mm *140mm Z-fold thermal paper
Patient current leakage <10μA Communication RS232/ USB port
Calibrating voltage 1mV±2% Power Supply AC 110-230V(±10%),50/60Hz(±1Hz),60VA

DC Built-in rechargeable Li-ion battery,14.4V (4400mAh)

Voltage tolerance ≥±500mV Dimension/Weight 347mm × 293mm × 83mm / 4.8Kg

Product Packing List

  • 1. Main Unit: 1 set
  • 2. Patient Cable: 1pc
  • 3. Chest Electrodes: 1set (6 pcs)
  • 4. Limb Electrodes: 1set (4 pcs)
  • 5. Power Cable: 1pc
  • 6. Grounding Cable: 1pc
  • 7. Fuse: 2pcs
  • 8. Paper Axis: 3pcs
  • 9. Thermal Paper: 1 roll
  • 10. Use Manual: 1 copy
  • 11. The factory Inspection Report: 1 copy