Quality Standards and Classification
CE, ISO13485
SFDA: Class Ⅱb
Anti-electroshock degree: ClassⅠequipment
(internal power supply)
TEMP/SpO 2 /NIBP: BF
ECG/Resp: CF
Application Range
Adult/ Pediatric /Neonatal/ Medicine/ Surgery/
Operating Room/ ICU/ CCU
Display
10”, 12.1”,15”real color TFT touch screen
Environment
Operating environment :
Temperature: 0 ~ 40 ℃
Humidity: ≤85%
Altitude: -500 ~ 4600m
Transport and Storage environment :
Temperature: -20 ~ 60 ℃
Humidity: ≤93%
Altitude: -500 ~ 13100m
Power Requirements
AC: 100 ~ 240V, 50Hz/60Hz
DC: Built-in rechargeable battery
Battery: 4000mA 11.1V lithium battery
2h operating after full charge(one piece)
5min operating after low battery alarm
Dimension and weight
Equipment:
E15: 360 mm × 162 mm ×321 mm; 4.47 kg
E15: 420 mm ×380 mm × 321 mm; 7.3 kg
Date Storage
Trend diagram/table: 7x24h
NIBP review: 400 events
Wave review: 100min
Alarm review: 100 alarm events
Support drug concentration titration analysis
ECG
5 Leads: RA, LA, LL, RL, V
Lead mode:Ⅰ,Ⅱ, Ⅲ, aVR, aVL, aVF, V
Increase: ×250,×500,×1000,×2000
Sensitivity: > 200 uV (Peak-to-peak value)
Input impedance: > 5 ( megohm )
Bandwidth: Surgery:1 ~ 20 Hz
Monitor: 0.5 ~ 40 Hz
Diagnostic: 0.05 ~ 130 Hz
CMRR: ≥ 100dB
Polarization Voltage: ± 300mV
Baseline Recovering Time: After defibrillation< 3 seconds
Signal Range: 8 mV(Peak-to-peak value)
Calibrating Signal: 1mV, precision ±5%
ST Segment
ST Segment Range: -2.0mV ~ +2.0mV
Accuracy: 0.02mV
RESP
Method: RA-LL impedance
Resp Impedance range: 0.3 ~ 3 Ω
Base Impedance range: 200 Ω ~ 4000 Ω
Bandwidth: 0.1 ~ 2.5Hz
Resp rate: Adult: 0 ~ 120BrPM
Neonatal/Pediatric: 0 ~ 150BrPM
Resolution: 1BrPM
Precision:±2 BrPM
Asphyxia Alarm: 10~40 seconds |
NIBP
Method: Pulse wave oscillometry
Work mode: Manual/ Auto/ STAT
Measure interval of auto mode:
1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)
Measuring Time of STAT Mode: 5 minutes
PR range: 40 ~ 240bpm
Measure & alarm range:
Adult: SYS 40 ~ 270mmHg
DIA 10 ~ 215mmHg
MEAN 20 ~ 235mmHg
Pediatric: SYS 40 ~ 200mmHg
DIA 10 ~ 150mmHg
MEAN 20 ~ 165mmHg
Neonatal: SYS 40 ~ 135mmHg
DIA 10 ~ 100mmHg
MEAN 20 ~ 110mmHg
Static pressure range: 0 ~ 300mmHg
Precision:± 3mmHg
Pressure precision:
Max. average error:±5mmHg
Max. standard deviation:±8mmHg
Overvoltage protection:
Adult: 300mmHg
Pediatric: 240mmHg
Neonatal: 150mmHg
SpO2
Range: 0 ~ 100% Resolution: 1%
Precision: 70% ~ 100%: ±2 DIGIT
0% ~ 69%:±no definition given
Pulse Rate
Range: 20 ~ 300bpm Resolution: 1bpm
Precision: ±3bpm
TEMP
Channel: 2
Measure & alarm range: 0 ~ 50 ℃
Resolution: 0.1 ℃
Precision (no sensor):±0.1 ℃
Standard Accessories
1. NIBP cuff & tube
2. ECG cable & electrodes
3. SpO2 sensor
4. TEMP probe
5. Lithium-ion battery
6. Power cable
7. Operator’s manual
Optional
1. CO2 module 2.IBP module 3.Trolley bracket
4. Hanging bracket 5.Monitor recorder
IBP
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART: 0 ~ 300 mmHg
PA: -6 ~ 120 mmHg
CVP/RAP/LAP/ICP: -10 ~ 40 mmHg
P1/P2: -10 ~ 300 mmHg
Press Sensor
Sensitivity: 5 uV/V/mmHg
Impedance: 300-3000Ω
Resolution: 1 mmHg
Accuracy: ±2% or ±1mmHg, which great
Actualization interval: about 1 Sec.
CO2
Side/Main stream
Measurement Infrared radiation absorption technique
measurement range: 0 ~ 99mmHg
Precision: 0 ~ 40mmHg ±2mmHg
41 ~76mmHg ±5mmHg
77~99mmHg ±10mmHg
Resolution: 1mmHg
Pumping rate: 70ml/min, 100ml/min
Accuracy: 15% or 15ml/min big
AwRR measuring range: 0 ~120BrPM
Precision: 0 ~70BrPM: ±2BrPM
>70BrPM: ±5BrPM |